We have used prethinned sapphire substrates to make a YBa2Cu3O7−x thin-film, transition-edge bolometer with peak detectivity of 6×109 cm Hz1/2/W near 4 Hz and 90 K. This is a new record for high Tc superconductor bolometers, and is comparable to the best detectivity yet achieved for a thermal-type infrared detector with no more than moderate cooling (liquid-nitrogen temperature and above). We discuss some simple improvements which may push the detectivity above 1010.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.