Abstract

Reinforcement of tissues by use of textiles is encouraged by the reduced rate of recurrent tissue dehiscence but for the price of an inflammatory and fibrotic tissue reaction to the implant. The latter mainly is affected by the size of the pores, whereas only sufficiently large pores are effective in preventing a complete scar entrapment. Comparing two different sling implants (TVT and SIS), which are used for the treatment of urinary incontinence, we can demonstrate that the measurement of the effective porosity reveals considerable differences in the textile construction. Furthermore the changes of porosity after application of a tensile load can indicate a structural instability, favouring pore collapse at stress and questioning the use for purposes that are not “tension-free.”

Highlights

  • Reinforcement of tissues by use of textile implants is increasingly used to improve the recurrence rates compared to unification of tissues just by sutures

  • The latter mainly is affected by the size of the pores, whereas only sufficiently large pores are effective in preventing a complete scar entrapment

  • Comparing two different sling implants (TVT and SIS), which are used for the treatment of urinary incontinence, we can demonstrate that the measurement of the effective porosity reveals considerable differences in the textile construction

Read more

Summary

Introduction

Reinforcement of tissues by use of textile implants is increasingly used to improve the recurrence rates compared to unification of tissues just by sutures. Comparing two different sling implants (TVT and SIS), which are used for the treatment of urinary incontinence, we can demonstrate that the measurement of the effective porosity reveals considerable differences in the textile construction.

Results
Conclusion