Abstract
A detailed procedure is described for the laboratory construction of electrodeless discharge lamps. Although this work pertains to Hg in Ar discharges, the same general experimental techniques are also applicable for other substances. The lamps made were without any getters and had an intensity variation of less than one part in 1000 over a period of 50 h of continuous operation. The described procedure allows for continuous monitoring of critical cleaning procedures in the evacuated lamp blank before it is filled and sealed off.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.