Abstract

Techniques for high-speed delay scanning are important for low-coherence interferometry, optical coherence tomography, pump probe measurements, and other applications. We demonstrate a novel scanning delay line using a multiple-pass cavity. Differential delays are accumulated with each pass so that millimeter delays can be generated with tens of micrometer mirror displacements. With special design criteria, misalignment sensitivity can be dramatically reduced. The system is demonstrated to scan 6 m/s at 2-kHz repetition rates. Real-time optical coherence tomography imaging with 500 pixel images at four frames/s is performed. Using a Cr:forsterite laser source, we obtained axial image resolutions of 6 microm with 92-dB sensitivity.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.