Abstract

We have developed a high speed instrument for automated DNA sequence analysis. The apparatus employs laser excitation and a cooled CCD detector for the parallel detection of up to 18 sets of four fluorescently labeled DNA sequencing reactions during their electrophoretic separation in ultrathin (50-100 microns) denaturing polyacrylamide gels. Four hundred and fifty bases of sequence information is obtained from 100 ng of M13 template DNA in less than one hour, corresponding to an overall instrument throughput of over 8000 bases/hr.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call