A high-speed highly sensitive and reproducible method for the determination of catecholamines in small plasma volumes has been developed. Catecholamines were separated on a short capillary fused-silica column (0.32 mm i.d.) packed with a reverse-phase sorbent with a particle size of 10 microns. The time of a single chromatographic analysis was less than 6 min, and the operating pressure in the system was not higher than 5.0 MPa. The required volume of blood plasma was 0.5 mL. When a 10 microL sample previously concentrated on alumina was injected into the column, the limits of amperometric detection for noradrenaline (NA), adrenaline (A), 3,4-dihydroxy-benzylamine (DHBA), and dopamine (DA) were 3.9, 2.6, 1.8 and 5.6 ng/L, respectively. The average variation coefficient was 17.4%.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.