Abstract

In this work, a finite element model was created using MRI scans of the main author to analyze sources of the dynamic thoracic bioimpedance. This model can be used to identify limitations of impedance cardiography (ICG) in practice. Heart beat (8.3 ms temporal resolution) and aortic wave propagation (2.6 ms temporal resolution) were implemented. The static volume contains all major organs of the thorax in high spatial resolution. Simulations were successfully conducted and a high correlation (r = 0.9) between the simulated aortic ICG signal and a measured signal of the same subject was obtained.