Abstract

High-sensitivity measurement of serum estrogens is important in adult and pediatric endocrinology and oncology. We developed a high-sensitivity liquid chromatography-tandem mass spectrometry (LC-MS/MS) assay for simultaneous measurement of estrone (E1) and estradiol (E2). Aliquots of 200 muL of serum were spiked with internal standard, extracted, derivatized with dansyl chloride, and analyzed by LC-MS/MS using 2-dimensional chromatographic separation. Total imprecision for the method was less than 11%; the limit of quantitation was 1 pg/mL. Reference intervals were established with samples from more than 900 healthy postmenopausal women, men, girls, and boys. Concentrations of estrogens in children reached adult levels by Tanner stage 3. In men and postmenopausal women, the median concentrations of total estrogens (E1 + E2) were 39 and 22 pg/mL, and the median E2/E1 ratios were 0.98 and 0.55, respectively. The method requires a small sample volume and has adequate sensitivity and specificity for analyzing estrogens in samples from postmenopausal women, men, and children.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.