Abstract

We demonstrate an ultrahigh sensitivity silicon photonic biosensor based on cascaded Mach-Zehnder interferometer (MZI) and ring resonator with the Vernier effect using wavelength interrogation. Experimental results show that the sensitivities reached 2870 nm/RIU and 21,500 nm/RIU for MZI sensor and MZI-ring sensor, respectively. A biosensing application was demonstrated by monitoring the interaction between goat and antigoat immunoglobulin G (IgG) pairs. The measured results show that 1 ng/ml IgG resulted in 0.035 nm and 0.5 nm wavelength shift for MZI sensor and MZI-ring sensor, respectively. This high performance sensor is promising for medical diagnostic applications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.