Abstract

The effect of acetylcholinesterase (AChE) immobilization over the surface of a quartz crystal microbalance (QCM), by chemisorption of the AChE thiolated with a heterobifunctional cross-linker, sulfo-succinimidyl-6-[3-(2-pyridyldithio)propionamido]hexanoate, and carboxyl-amine coupling of AChE to 3-mercaptopropionic acid self-assembled monolayer, on the responses of a batch-type QCM-precipitation sensor was compared, resulting in a better sensitivity and binding efficiency in the former method. When an inhibition study with the developed sensor was undertaken at the optimized AChE immobilization with varying concentrations of a model organophosphorus pesticide EPN and carbamate one carbofuran, a sensitive detection for them was possible with the limit of detection corresponding to 1.55 x 10(-8) and 1.30 x 10(-9)M, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.