Abstract

Raman spectroscopy with the advantages of the in situ and simultaneous detection of multi-components has been widely used in the identification and quantitative detection of gas. As a type of scattering spectroscopy, the detection sensitivity of Raman spectroscopy is relatively lower, mainly due to the low signal collection efficiency. This paper presents the design and assembly of a multi-channel cavity-enhanced Raman spectroscopy system, optimizing the structure of the sample pool to reduce the loss of the laser and increase the excitation intensity of the Raman signals. Moreover, three channels are used to collect Raman signals to increase the signal collection efficiency for improving the detection sensitivity. The results showed that the limits of detection for the CH4, H2, CO2, O2, and N2 gases were calculated to be 3.1, 34.9, 17.9, 27, and 35.2 ppm, respectively. The established calibration curves showed that the correlation coefficients were all greater than 0.999, indicating an excellent linear correlation and high level of reliability. Meanwhile, under long-time integration detection, the Raman signals of CH4, H2, and CO2 could be clearly distinguished at the concentrations of 10, 10, and 50 ppm, respectively. The results indicated that the designed Raman system possesses broad application prospects in complex field environments.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.