Abstract

The incoherent Thomson scattering diagnostic with multiple lines of sight is installed at the gas dynamic trap (GDT) for measurements of radial profiles of the plasma electron temperature and density. The diagnostic is built on the Nd:YAG laser operating at 1064nm. The laser input beamline is provided with an automatic system for alignment status monitoring and correction. The collecting lens uses ∼90° scattering geometry having 11 lines of sight in total. Presently, six of them (covering the full plasma radius from the axis to the limiter) are equipped with high etendue (f/2.4) interference filter spectrometers. The design of the spectrometer's data acquisition system based on the "time stretch" principle allowed for the 12 bits vertical resolution with a sampling rate of 5 GSample/s and a maximum sustainable measurement repetition frequency of 40kHz. The repetition frequency is the crucial parameter for the study of plasma dynamics with a new pulse burst laser to be started in early 2023. Results of the diagnostic operation in several GDT campaigns show that radial profiles are routinely delivered with the typical observation error of 2%-3% for Te ⩾ 20eV in a single pulse. After Raman scattering calibration, the diagnostic is capable to measure the electron density profile with the resolution ne (min)≃4⋅1018m-3 and error bars of 5%.

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call