Abstract

Based on work by Dubochet and others in the 1980s and 1990s, samples for single-particle cryo-electron microscopy (cryo-EM) have been vitrified using ethane, propane or ethane/propane mixtures. These liquid cryogens have a large difference between their melting and boiling temperatures and so can absorb substantial heat without formation of an insulating vapor layer adjacent to a cooling sample. However, ethane and propane are flammable, they must be liquified in liquid nitro-gen immediately before cryo-EM sample preparation, and cryocooled samples must be transferred to liquid nitro-gen for storage, complicating workflows and increasing the chance of sample damage during handling. Experiments over the last 15 years have shown that cooling rates required to vitrify pure water are only ∼250 000 K s-1, at the low end of earlier estimates, and that the dominant factor that has limited cooling rates of small samples in liquid nitro-gen is sample precooling in cold gas present above the liquid cryogen surface, not the Leidenfrost effect. Using an automated cryocooling instrument developed for cryocrystallography that combines high plunge speeds with efficient removal of cold gas, we show that single-particle cryo-EM samples on commercial grids can be routinely vitrified using only boiling nitro-gen and obtain apoferritin datasets and refined structures with 2.65 Å resolution. The use of liquid nitro-gen as the primary coolant may allow manual and automated workflows to be simplified and may reduce sample stresses that contribute to beam-induced motion.

Highlights

  • Single-particle cryo-electron microscopy (Frank, 2002) has emerged as a powerful approach to obtaining near atomic resolution structures of large biomolecular complexes, membrane proteins, and other targets of major scientific, pharmaceutical and biotechnological interest (Cheng, 2015, 2018; Glaeser, 2016b, 2019; Vinothkumar & Henderson, 2016; Lyumkis, 2019)

  • Unlike X-ray crystallography, cryo-electron microscopy (cryo-EM) requires only a small amount of biomolecular sample dispersed in solution

  • As in cryo-crystallography, the key challenges in singleparticle cryo-EM are associated with sample preparation and handling

Read more

Summary

Introduction

Single-particle cryo-electron microscopy (cryo-EM) (Frank, 2002) has emerged as a powerful approach to obtaining near atomic resolution structures of large biomolecular complexes, membrane proteins, and other targets of major scientific, pharmaceutical and biotechnological interest (Cheng, 2015, 2018; Glaeser, 2016b, 2019; Vinothkumar & Henderson, 2016; Lyumkis, 2019). Unlike X-ray crystallography, cryo-EM requires only a small amount of biomolecular sample dispersed in solution. It allows the structural study of systems that have been intractable to crystallization and is becoming a go-to method for initial attempts at structure determination. To obtain vitrified buffer for the best imaging, the sample-containing grid is plunged at 1–2 m sÀ1 into liquid ethane at T ’ 90 K (produced by cooling ethane gas in a liquid-nitrogen-cooled cup). The sample is transferred from ethane to liquid nitrogen (LN2), loaded into grid boxes, transferred to additional containers and to a storage Dewar. Samples are removed from the storage Dewar and grid boxes, and loaded into a cold microscope stage. The stage or cassette is loaded into the TEM

Methods
Results
Discussion
Conclusion
Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.