Abstract

High-resolution three-headed single photon emission computed tomography (SPET) equipped with fan-beam collimators was applied to myocardial perfusion imaging in infants aged from 1 to 11 months (n = 5). A tabletop designed specifically for infants was fixed on the SPET couch to reduce the radius of camera rotation to 13.2 cm. Significant improvement in resolution was achieved with the fan-beam collimators compared to parallel-hole high-resolution collimators. With the administration of approximately 37 MBq (26-44 MBq) 201Tl, 5 min acquisition time was possible for SPET imaging, which provided good image quality in all patients. Thus, a smaller administration dose is possible within a practical short acquisition time. High-resolution fan-beam SPET imaging can be a routine diagnostic method for heart disease in newborn babies and infants.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.