Abstract

This paper presents preliminary results of research toward the development of a high resolution analysis stage for a dual resolution image processing-based prescreening device for cervical cytology. Experiments using both manual and automatic methods for cell segmentation are described. In both cases, 1500 cervical cells were analyzed and classified as normal or abnormal (dysplastic or malignant) using a minimum Mahalanobis distance classifier with eight subclasses of normal cells, and five subclasses of abnormal cells. With manual segmentation, false positive and false negative error rates of 2.98 and 7.73% were obtained. Similar experiments using automatic cell segmentation methods yielded false positive and false negative error rates of 3.90 and 11.56%, respectively. In both cases, independent training and testing data were used.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.