Abstract

AbstractIntroductionChoosing a hinged implant in the revision knee arthroplasty (rTKA) setting is challenging and limited data on implant performance exists. We present the survivorship and reason for failure in rTKA performed at our institution using the LINK hinge prosthesis, predominantly the cemented modular Endo-Model prosthesis.Methodology260 consecutive revision knee cases performed between 2012 and 2020 were reviewed retrospectively. Mean follow up was 27 months (range 0 to 107). Survivorship was analysed in Stata using a Log Rank test to compare performance in patients stratified according to age (≥80 years old (76 cases), 70–79 years (104 cases) and ≤70 years (80 cases).Results53 patients died and 48/207 (23%) cases in 40 patients underwent re-revision. Reasons for re-revision were aseptic loosening (21), infection (12), instability (4), extensor failure (1), stiffness (1), fracture (1) and other (8). Loosening was seen in the femur (8), tibia (5), and both the femur and tibia (8). Sub-group analysis of patients according to age showed a significantly higher failure rate in younger patients (6 failures (8%) in patients ≥80, 27 failures (26%) in 70–79 and 15 (19%) in ≤70 (p = 0.02). Failure in patients ≤70 was predominantly due to aseptic loosening (8/15).ConclusionHere we report a significantly higher rate of LINK hinge prostheses failure in patients <70 undergoing rTKA. Consent should consider the risk of re-revision in this patient group.