Abstract

The availability of germanium with less than 1010 acceptors/cm3 has enabled us to fabricate and study the performance of germanium detectors of substantially larger volume than has hitherto been possible except by using lithium-drifting techniques. Detectors having a sensitive volume of up to 25 cm3 have been made. These detectors have a lithium diffused n+ contact and a metal barrier non-injecting back contact, and are totally depleted. The energy resolution of these detectors is similar to that of equivalent sized lithium-drifted detectors, but the high-purity detectors are easier to fabricate, and present considerably fewer handling problems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.