A hydrostatic high-pressure cell (maximum pressure 700 MPa) with synthetic diamond windows is applied to measure small-angle X-ray scattering of a protein at high pressure. Use of the present cell allows an accurate solvent background correction, providing quantitative analyses. The performance of the present cell for X-ray scattering is shown by using lysozyme as a sample solution. From the Guinier plot, values of the radius of gyration of lysozyme are evaluated to be 15.31 +/- 0.09 A at 1 atm (0.10 MPa) and 14.80 +/- 0.15 A at 400 MPa.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.