Abstract

A new sapphire capillary pressure cell for single-crystal X-ray diffraction measurements at moderate pressures (200−1500 bar; 1 bar = 100 kPa) has been developed and optimized for use on beamline I19 at Diamond Light Source. The three-component cell permits optical centring of the crystal andin situpressure modification to a precision of 1 bar. Compression of hexamethylenetetramine and its deuterated analogue to 1000 bar was performed, showcasing the accuracy and precision of the measurements, and highlighting evidence of a geometric isotope effect.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.