Abstract

A high pressure liquid chromatographic method has been developed for determining aflatoxins B1, B2, G1, and G2 in peanut butter. The method is based on extraction with acidified aqueous methanol, partition of the aflatoxin into methylene chloride, and purification of the extract on a 2 g silica gel column. The extracted aflatoxins are resolved on a microparticulate (10 micrometer) porous silica gel column in ca 10 min with a water-washed chloroform-cyclohexane-acetonitrile solvent that contains 2% isopropanol. The fluorescence detection system determines aflatoxins B1, B2, G1, and G2 at low levels, i.e., 0.25 ppb B1, 0.5 ppb G1, and 0.2 ppb B2 and G2. Multiple assays of 5 samples of naturally contaminated peanut butters containing total aflatoxins (B1 + B2 + G1 + G2) at levels of 1, 2, 3, 9, and 17 ppb gave intralaboratory coefficients of variation of 7, 4, 4, 11, and 3%, respectively. Samples spiked at levels of 5, 9, and 17 ppb total aflatoxins showed recoveries of 79, 81, and 81%, respectively.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.