Abstract
In the present work, design and operation of a high pressure gas adsorption apparatus at room temperature and at pressures up to 100 bar are discussed. A theoretical and experimental error analysis is done to determine accuracy and robustness of the measurements. For this study, activated carbon was selected as the adsorbent and hydrogen as the adsorbate gas. A sensitivity analysis was done by taking into account the effects of temperature, pressure, volume and weight of the sample. The analysis shows that the volumes of the sample and reference cells as determined by helium-free space measurements have significant effect on the accuracy of the adsorption uptake measurement. For instance, a 0.1% error in the measurement of either volume led to approximately a 3% error in hydrogen uptake measurement at 298 K and 100 bar.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.