Abstract

Oxygenator high pressure excursion (HPE) is a phenomenon that can occur during cardiopulmonary bypass (CPB) in which the oxygenator inlet pressure increases rapidly, thereby limiting flow. Most perfusionists globally do not use inlet oxygenator pressure monitoring and therefore HPE is not often recognized. HPE may occur for various reasons, and it is not fully understood. Patient factors that put a patient at a higher risk of HPE are increased body surface area, blood type, and hematocrit count. Size, blood flow, and pressure drops of the oxygenator incorporated into the circuit can also increase the probability of an HPE occurring. This case study overviews our experience when dealing with an interesting case of HPE and the most up to date knowledge on appropriate steps to mitigate the effects on the patient.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.