Abstract

Application of piezoelectric flexural mechanical resonators such as tuning forks to accurate measurements of liquid physical properties is discussed. It was shown earlier that liquid properties such as viscosity, density and dielectric constant can be obtained by measuring the resonator AC impedance within certain frequency range and fitting it to the resonator equivalent circuit model (1). Error sources for the liquid property measurements and their influence on the measured value are investigated. It is shown experimentally that the reproducibility of the viscosity and density measurements using this technique can meet and often exceed the one delivered by the well established analytical instrumentation. It is also demonstrated here that better performance is resulting from the use of the whole impedance curve over a frequency range, which produces better statistics and natural averaging of the noise.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.