Abstract
A high-precision phase measurement technique, which can accurately determine the phase even when the signals are distorted by harmonics, was presented previously by the author (ibid., vol.38, p.954-60, Oct. 1989). He proposes a microprocessor-based hardware implementation of the technique. The hardware implementation requires two reduced tables: a sine table and a cosine table. If each table has i(i+5)/2 entries, then the phase can be accurately determined even if the signals are distorted by any number of harmonics in the range of second through ith harmonics.< <ETX xmlns:mml="http://www.w3.org/1998/Math/MathML" xmlns:xlink="http://www.w3.org/1999/xlink">></ETX>
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: IEEE Transactions on Instrumentation and Measurement
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.