Abstract

Objective To research and design a kind of high accuracy laser power density meter to measure laser power density in clinical medicine,in order to combine laser with clinical medicine well.Methods The pyroelectric detector was used to convert the laser signal to available electrical signal,preamplifier filter circuit and control circuit were designed to match the detector,and C was used for software programming.With the combination of hardware and software design,a high precision laser power density meter was developed,which was based on AVR microcontroller.Results The instrument could measure the minimum 1 mW/cm2 of laser power density accurately and the error rate was 2%,which met the requirements of the laser power density meter parameters in medical application.Conclusion The power density meter has a high precision,small error rate and good stability,and it can measure the laser power density accurately. Key words: Pyroelectric detector; Laser power density meter; High-precision measurement; AVR microcontroller

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.