Abstract

We are completing the construction of GIANO, a high resolution near-infrared cryogenic spectrograph for the Telescopio Nazionale Galileo (TNG). Most of the optics are made of aluminium and operate at cryogenic temperature. We evaluated the optical degradation due to mis-matches between the thermal expansion coefficients of the different aluminium parts of the instrument. We performed accurate measurements of the relative thermal expansion coefficients (CTE) of Al-6061 and Al-6082 over the 300–77 K temperatures range. We find that the two alloys have identical thermal expansion coefficient within a maximum (3σ) uncertainty of Δα/α < 0.28%. Our results show that it is possible to overcome the problem of the alignment of a cryogenic instrument, manufacturing the curved optics, the optics’ holders and the optical bench with different metallic alloys with small CTE mismatch (Al-6061 and Al-6082). This conclusion has also been confirmed by the results of the optical tests with the instrument cooled in the laboratory, showing no significant image quality degradation.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.