Abstract

Here we describe a new analytical technique for the high-precision measurement of 182W/184W and 183W/184W using negative thermal ionization mass spectrometry (N-TIMS). We improve on the recently reported method of Trinquier et al. (2016), which described using Faraday cup collectors coupled with amplifiers utilizing 1013Ω resistors to continuously monitor the 18O/16O of WO3− and make per-integration oxide corrections. In our study, we report and utilize a newly measured oxygen mass fractionation line, as well as average 17O/16O and 18O/16O, which allow for more accurate per-integration oxide interference corrections. We also report a Faraday cup and amplifier configuration that allows 18O/16O to be continuously monitored for WO3− and ReO3−, both of which are ionized during analyses of W using Re ribbon. The long-term external precision of 182W/184W is 5.7ppm and 3.7ppm (2SD) when mass bias corrected using 186W/184W and 186W/183W, respectively. For 183W/184W mass bias is corrected using 186W/184W, yielding a long-term external precision of 6.6ppm. An observed, correlated variation in 182W/184W and 183W/184W, when mass bias corrected using 186W/184W, is most likely the result of Faraday cup degradation over months-long intervals.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.