Abstract

A high-power argon laser system (maximum 20 W on the skin) was used with a 2-mm-spot size for the treatment of port-wine stains (PWS). Five hundred and thirty test treatments were evaluated. With short pulse times (< 15ms), relatively specific damage to vessels could be demonstrated histologically, but sufficient clinical lightening of PWS was seen only in 30% of patients due to the resulting low energy fluence (∼=10 Jcm−2_. With higher energy fluences and longer pulse times, clinical lightening increased but vessel specificity was lost. Overall results showed that with a high-power argon laser comparable clinical blanching can be achieved as with conventional argon lasers, but the necessary energy fluences are lower (∼=15 Jcm−2) leading to less pain under treatment and a minimal incidence of side-effects.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.