Abstract

Through computer simulations and surface plasmon resonance (SPR) measurements, we establish optimum parameters for the design and fabrication of SPR sensors of high sensitivity, resolution, stability, and long decay-length evanescent fields. We present simulations and experimental SPR data for variety of sensors fabricated by using bimetal (Ag/Au) and multilayer waveguide-coupled Ag/Si3N4/Au structures. The simulations were carried out by using the transfer matrix method in MATLAB environment. Results are presented as functions of the thickness of the metal (Ag or Au) and the waveguide dielectric used in Ag/Si3N4/Au structures. Excellent agreement is observed between the simulations and experiments. For optimized thickness of the Si3N4 waveguide (150 nm), the sensor exhibits very high sensitivity to changes in the refractive index of analytes, Sn≈52°/RIU, extremely high resolution (FWHM≤0.28°), and long penetration depth of evanescent fields (δ≥305nm).

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.