Abstract

ADVIA Centaur SARS-CoV-2 Antigen (COV2Ag) Assay (Siemens Healthineers) was evaluated for SARS-CoV-2 detection. A total of 141 nasopharyngeal samples were analyzed by this technique and results were compared with those obtained by quantitative reverse-transcription polymerase chain reaction (RT-PCR). The overall sensitivity and specificity of the test were 68.70% and 70%, respectively. Regarding cycle threshold (Ct) values, the COV2Ag test showed a sensitivity of 93.75% and 100% for nasopharyngeal samples with Ct < 25 and < 20, respectively. ADVIA Centaur COV2Ag Assay is a useful, automated, and rapid technique for early SARS-CoV-2 diagnosis and isolation of the infected individuals, avoiding its transmission.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.