Abstract

High-performance liquid chromatography (h.p.l.c.) is used for separation of arsenite, arsenate, monomethylarsinate (MMA) and dimethylarsonate (DMA) followed by continuous sodium tetrahydroborate reduction and atomic absorption spectrometric detection. Sample preconcentration, offering improved detection limits for the individual species and the removal of matrix interferences, is achieved with a pellicular anion-exchange column. The arsenic species are then separated on a strong anion-exchange column placed in series with the preconcentration column. Detection limits of 2 ng (as arsenic) for arsenite, arsenate and MMA, and 1 ng for DMA. Results for arsenic species in soil waters and commercial bottle waters are given.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.