Abstract

This report describes a specific and precise high-performance liquid chromatography (HPLC) method for the quantification of trans,trans-muconic acid in human urine. The procedure involved a highly efficient Bond-Elut SAX extraction with 20% acetic acid elution. The HPLC analysis used a sodium acetate/methanol mobile phase with a C18 reverse phase column and UV detection at 265 nm. The recovery, precision, linearity, and limits of detection and quantification of the method were determined. Mean absolute recoveries were between 97% and 115%. The calibration curve showed a correlation coefficient of 0.9955 and the limit of detection was determined to be 10.8 microg/L. The method is suitable for evaluation of occupational and environmental benzene exposure in humans. The study of urinary trans,trans-muconic acid of two populations of children to evaluate environmental benzene exposure is presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.