Abstract

Therapeutic drug monitoring (TDM) and dose adjustment of lenvatinib may be beneficial in the treatment of radioiodine-refractory thyroid cancer, by maximizing antitumor effects and minimizing adverse drug reactions. The aim of this study was, therefore, to develop and validate a high-performance liquid chromatography method using an ultraviolet detection system for routine serum lenvatinib detection in patients with thyroid cancer. Serum specimens, spiked with an internal standard, were treated by a solid-phase extraction through an octadecylsilyl silica cartridge. Lenvatinib and internal standard were concomitantly separated from serum using a conventional octadecylsilyl silica column through isocratic elution, using a mobile phase consisting of 0.02 mol/L sodium phosphate (pH 6.7) and acetonitrile (50/50, vol/vol) at a flow rate of 1.0 mL/min. The detection wavelength was set at 244 nm. Serum samples from 5 patients were used for clinical validation of the method. The calibration curve for lenvatinib was linear (Pearson correlation coefficient, r = 0.9998) over the concentration range of 6.25-400 ng/mL, with a lower limit of quantification of 6.25 ng/mL. Extraction recoveries for lenvatinib were 97% or more, with coefficients of variation less than 2.2%. The coefficients of variation for intraday and interday assays were less than 4.7% and 6.0%, respectively. This sensitive high-performance liquid chromatography method can be used for lenvatinib therapeutic drug monitoring when liquid chromatography-tandem mass spectrometry facilities are unavailable.