Abstract
A simple method for the determination of α-tocopherol in vegetables is described. The procedure consists of the following steps: saponification, extraction, silica-column clean-up, and high-performance liquid chromatography. Elution time for D, L-α-tocopherol was 9.0 min using a Zorbax Sil (250 x 4.6 mm) column and an isocratic mobile phase of hexane-methanol (99.3 + 0.7), with a flow rate of 1 ml/min, and detection at 292 nm using a variable UV detector. The average recovery of α-tocopherol was 91.2%, and the minimum detectable amount was 0.1 mg/100 g of fresh vegetable tissue. This method is comparable to gas-chromatographic determination of α-tocopherol, but has fewer analytical steps and gives more reproducible results.
Highlights
Summary
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.