Abstract

A method for the determination of doxorubicin and daunorubicin in plasma is described. The plasma is injected directly into a loop column and then washed with water. After switching the injection valve, the sample is separated on a phenyl column using detection at 254 nm. The detection limit is 10 ng/mL, the coefficient of variation is 7% for 100 ng/mL of doxorubicin and 4% for 200 ng/mL of daunorubicin.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.