Abstract

AbstractHPLC and CE methods were employed to determine the quaternary benzo[c]phenanthridine alkaloid sanguinarine in biological matrices (rat hepatocytes, human gingival fibroblasts, feed, porcine faeces, body fluids and tissues). HPLC was carried out on a C18 column using gradient elution and ion pairing techniques with 1‐heptanesulfonic acid as ion pairing agent under acidic conditions. The detection limit for fluorimetric detection at λex = 327 nm and λem = 577 nm was 3 nM sanguinarine. CE analyses were performed in 50 mM phosphate‐Na buffer pH 2.5, with 150 mM SDS used for pre‐concentration by the sweeping effect. This experimental configuration allows injecting the total capillary length with sanguinarine sample. The detection limit for UV detection at 285 nm was 12 nM. Both methods are suitable for analysing submicromolar quantities of sanguinarine in biological materials. The HPLC method is more sensitive than CE because it uses fluorescence detection.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.