Abstract

The reagent bis(acetylpivalylmethane)ethylenediimine has been examined for the HPLC separation of copper(II), nickel(II), palladium(II) and oxovanadium(IV) chelates on reversed phase HPLC columns (250 × 4 mm) packed with Hypersil ODS, 5 μ and (150 × 3.9 mm) Nova Pak C 18 with guard column. The complexes are eluted with a binary mixture of methanol and water or methanol, acetonitrile and water. Detection is achieved with a UV detector. The solvent extraction procedure is used for the determination of copper and nickel simultaneously at microgram levels, corresponding to ng levels and pg levels respectively, per injection. The method has been applied to the determination of copper and nickel in a coin, nickel—aluminum alloy and water samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.