Abstract

A method is described for the determination of bradykinin, neuropeptide K (NPK), and substance P in patients with atypical carcinoid syndrome. The developed method uses a combination of conventional and solid-phase extraction as well as high-performance liquid chromatographic techniques. A narrow-bore C18 column with ultraviolet detection is used (200 nm). The technique recovers bradykinin at a level of 98%, NPK at 96%, and substance P at 98% (when pure standards are dissolved) at concentration levels relevant to the atypical carcinoid syndrome. In biological samples, the recovery rate of bradykinin, NPK, and substance P drops to 88, 86, and 88% respectively. The overall analysis time is 150 min from receipt of samples. This method proves to be a valuable tool in the identification of neuropeptides and thus the diagnosis of atypical carcinoid syndrome, especially in puzzling cases with nonspecific symptoms.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.