Abstract

A rapid, selective, and simple high-performance liquid chromatographic assay for digitoxin formulations is described. The method utilizes a conventional octadecyl-bonded phase column with detection at 220nm. The isocratic solvent system resolves digitoxin from its potential degradation products and provides an accurate assay for tablet and injectable formulations with a relative standard deviation of 1.4 and 3.3%, respectively. The method is sufficiently sensitive to monitor content uniformity of tablets and the minimum quantifiable amount of digitoxin was determined to be 20ng. The total chromatograph time was ~15min.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.