Abstract

Magnetic resonance imaging (MRI) guided intervention has drawn increasing attention over the last decade. It is accredited to the capability of monitoring any physiological change of soft tissue with the high-contrast MR images. This also gives rise to the demand for precise tele-manipulation of interventional instruments. However, there is still lack of choices of MR safe actuators that provide high-fidelity robot manipulation. In this letter, we present a three-cylinder hydraulic motor using rolling-diaphragm-sealed cylinders, which can provide continuous bidirectional rotation with unlimited range. Both kinematics and dynamics models of the presented motor were studied, which facilitate its overall design optimization and position/torque control. Motor performance, such as step response, frequency response, and accuracy, were experimentally evaluated. We also integrate the motor into our catheter robot prototype designed for intra-operative MRI-guided cardiac electrophysiology (EP), which can provide full degree-of-freedom and precise manipulation of a standard EP catheter.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.