Abstract

A description is given of pulsed gradient spin echo (PGSE) NMR experiments in which large pulsed magnetic field gradients may be required. The design contraints are discussed and, in particular, the problem of the use of large pulsed magnetic field gradients in conjunction with large polarizing fields is considered. Issues addressed concern probe mechanical assembly, current supply requirements, and pulse shape design. We describe a quadrupolar coil with a gradient amplitude of 1.65 T m−1A−1which has been used successfully up to a maximum gradient of around 40 T m−1. A diffusion coefficent of 7.5 × 10−16m2s−1has been measured using this system, the lowest yet achieved by PGSE NMR methods.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.