Abstract

We present a general analytical method of correcting for magnetic-field-induced changes in both the apparent temperature and the sensitivity of resistive thermometers at dilution refrigerator temperatures. With this method, we are able to reduce magnetic field induced errors in temperature to a level limited only by our ability to regulate the temperature in the absence of a magnetic field. We illustrate the application of our method to two resistive sensors in magnetic fields up to 18 T: a custom-made AuxGe1−x thin film sensor used in calorimetry and a commercially available ruthenium–oxide thick film resistor used in thermometry.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.