Abstract

Infection and asphyxia are two major causes of neonatal death globally. Where single-use resuscitation devices or sterilization of re-usable devices are unavailable, there is a need for effective, low-cost methods of high-level disinfection. Laboratory validation examined the efficacy of boiling and enclosed steaming (without pressure) as methods for attaining high-level disinfection of re-usable neonatal resuscitation equipment. The microbial load extracted and measured for each test article met internationally accepted standards for high-level disinfection. Boiling and steaming are low-cost, effective methods for reprocessing re-usable neonatal resuscitation devices in low- and middle-income countries.