Abstract

Thin silicon detectors have been used to measure high LET doses in patients being treated with negative pions. Measurements were made on the skin, in the oral cavity, and in the rectum of a series of patients and of certain selected patients. An analysis is given of the LET distribution of the particles observed in the thin, flat detector, and the high LET dose fraction is calculated from the event size spectrum. The data provide verification of the doses calculated from treatment planning. By timing the occurrence of the peak in the high LET dose distribution with respect to the position of the rangeshifter, one obtains the water-equivalent thickness of the bolus and tissues in front of the detector. This is the complementary measurement to the high LET dose measurement.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.