Abstract
To assess the effect of high intensity ultrasound energy on atherosclerotic arteries, the pressure-volume relation of stenoses in 14 atherosclerotic human cadaver arteries was measured before and after intraarterial application of ultrasound energy. The pressure-volume relation was measured by inflating a 3-mm angioplasty balloon within each artery with use of a syringe equipped with pressure and volume transducers. To minimize potential effects of balloon inflations on compliance, balloon inflation pressure was set at ≤3 atm before ultrasound application. Ultrasound energy was applied by using a titanium wire probe at a frequency of 20 kHz for 2 min.After exposure to ultrasound, arterial distensibility (measured as volume at 1 atm of balloon pressure) increased by a mean value of 82 ± 60 μl. These data suggest that ultrasound energy probably increases the pliability of atherosclerotic lesions by interrupting calcified plaque. This effect may enhance in vivo distensibility and render calcified atherosclerotic lesions more amenable to balloon angioplasty.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.