Abstract

AbstractThe use of secondary targets with commercially available high power X‐ray spectroscopy tubes has been investigated as a means of providing an excitation source for X‐ray energy spectroscopy. A comparison has been made between the secondary target mode of excitation and the use of the tube directly. Results from this comparison show the superiority of the secondary target technique. Examples of analysis using the secondary target excitation are given to illustrate the high data rates and low background this technique has to offer.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.