This paper deals with the radiative calibration aspects of high-heat-flux sensors using black-body radiation. In the last two years, several heat-flux sensors were calibrated up to 50 kW/m2 using a 25 mm diameter aperture variable-temperature black body and a reference room-temperature electrical-substitution radiometer. Tests on a typical Schmidt-Boelter heat-flux sensor showed long-term repeatability of calibration is within 0.6%. Plans for extending the present calibration capability to 100 kW/m2 are discussed.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.