Abstract

We present experimental results obtained at a user-oriented XUV beamline implemented at the ELI Beamlines facility. The coherent XUV radiation is produced via high harmonic generation in gases in a loose focusing geometry. The beamline is designed to be driven by 1 kHz, 100 mJ, 20 fs pulses centered at a wavelength of 830 nm. Results such as XUV spectra, beam wavefront and pulse energy obtained during the beamline commissioning with a commercial 1 kHz, 5 mJ, 40 fs laser system are presented. A unique XUV spectrometer for source characterization designed to reach a very high sensitivity is described in detail, and we demonstrate a novel technique for single-shot and every-shot XUV pulse energy measurement.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.