Abstract

Nuclear magnetic resonance (NMR) measurements using a newly designed hybrid magnet system installed at the National Institute for Materials Science were performed up to 28 T. A modified resistive insert magnet improved the field homogeneity of the hybrid magnet from 186 ppm/±5 mm to 16 ppm/±5 mm along the z-axis. Reconstruction of the power source for the resistive magnet suppressed the field instability from 30 to 3 ppmrms. These improvements enable us to obtain an NMR spectrum with narrower linewidth.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.