Abstract

We construct a photon-counting, fluorescence lifetime measurement system with a channel width of 5.0 ps. This is a modified version of our previous fluorometer based on a simultaneous detection method of photoelectron pulse trains after pulsed excitation (Iwata in Meas Sci Technol 28:075501, 2017), which can accept multiple photons during a one period of the excitation for building a histogram. Although the resolution time is 2.0 ns as before, the channel width of 5.0 ps is of the same order as that of time-correlated single-photon counting, which is achieved by delaying the trigger timing for the fluorometer by a step to that of the excitation. This results in an increase in the precision for the lifetime estimation. Although the measurement time increases with increasing the number of delay steps, it is still rather shortened owing to the high signal-gathering and histogram-building efficiencies.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.